3
Components Installation
Installing Valves and Fittings in the Unheated Compartment
58
TRACE 1310 Auxiliary Oven Instruction Manual
Figure 83. Unheated Rotary Valve Installation (5)
4. Connect gas lines from the solenoid valve manifold or the actuator.
Note Gas lines are much easier to connect to the actuator if the connection is made
before the actuator is assembled to the valve.
Figure 84. Unheated Rotary Valve Installation (4)
Actuator upper fitting
Actuator lower fitting
With reference to
Figure 84
a. Locate the nylon tubing supplied in the TRACE 1310 Auxiliary Oven kit and cut it
into two equal 12-in. pieces.
b. Push one of the tube pieces into the lower fitting on the actuator assembly, making
sure it passes through the ferrule in the actuator fitting. Tighten the nut and ferrule
using a 3/8-in. open end wrench. Use a backup 3/8-in. open end wrench on the
mating piece that is screwed into the actuator.
c. Place the other end of the tubing into the valve solenoid fitting located on the outside
edge of the solenoid manifold by pushing the tubing firmly into the fitting. It should
automatically seal in place.
Events Connector J24