Note Installation of the splitter:
The previous figures shown an example of installation of the splitter into the main
oven. You can place the splitter in a position that you feel more convenient based on
the system configuration. If you want to install the splitter on a wall of the main oven
as shown in the figures, proceed as follows:
Screw the dowel provided into the central hole of the splitter using a T8 Torxhead
wrench. Stop the screwing of the dowel when it starts to grip the mixer; the dowel
must protrude few millimeters from the central hole of the splitter, do not overtighten.
Screw the splitter into a hole provided on the wall of the main oven
Connect the splitter to the pneumatic circuit according to the scheme on
CAUTION To install the capillary column to the splitter, replace the steel ferrule inside the
splitter with the appropriate Vespel® ferrule provided with the kit.
Splitter
Dowel
From the port 6 of the Sampling Valve
Charcoal Trap
Analytical Column
TRACE 1310 Auxiliary Oven Instruction Manual