Thermo Scientific TRACE 1300 Instruction Manual page 131

Auxiliary ovens
Hide thumbs Also See for TRACE 1300:
Table of Contents

Advertisement

Thermo Scientific
2. Insert inner tube into the GC oven.
CAUTION Before proceeding, ensure that:
• The inner tube to the GC oven has been inserted into the passing hole provided
in the wall of the main oven but not completely fixed. See
• The biggest disks of insulating material are placed on the inner tube protruding
from the wall of the main oven. See
3. Carefully couple the TRACE 1310 Auxiliary Oven as a right panel ensuring that the
inner tube assembly penetrates into the right wall of the GC through the duct previously
prepared. See
Figure
157.
Figure 157. Coupling Valve Oven Module and GC (1)
a.
4. From the interior of the GC oven, place the smallest insulating disk on the inner tube,
then push the insulating disk into the duct. See
Figure 158. Coupling Valve Oven Module and GC (2)
3
Coupling the TRACE 1310 Auxiliary Oven to the GC
Figure 9
Figure 10
on
page
16.
Figure
158.
Smallest Insulating Disk
TRACE 1310 Auxiliary Oven Instruction Manual
Components Installation
on
page
15.
105