Wiring the TRACE 1310 Auxiliary Oven to the GC

Thermo Scientific
3. Guide the transfer line into the main oven through a convenient hole on the wall of the
main oven. See
Figure
Figure 139. Installing a Heated Sample Transfer Line (3)
This section provides the instructions for wiring the TRACE 1310 Auxiliary Oven with the
GC.
To wire the Valve Oven to the GC
CAUTION Before starting ensure that the GC is powered off and the power cable is
unplugged from the AC Input connector into the back of the GC and from the wall
outlet.
1. Remove the top cover of the GC.
a. Using a T20 Torxhead screwdriver, unscrew the four screws that held it in place.
See
Figure 140
and
Figure 140. Top Cover Removal (1)
b. Lift the top panel up and off the GC.
139.
Figure
141.
TRACE 1310 Auxiliary Oven Instruction Manual
3
Components Installation
Wiring the TRACE 1310 Auxiliary Oven to the GC
Top Cover Screws
95