Installing an Unheated Rotary Valve into the Unheated Compartment

Thermo Scientific
4. Reinstall the plate on the frontal wall of the unheated compartment.
a. Using a cross-head screwdriver screw the four fixing screws of the plate.
See
Figure
78.
Figure 78. Needle Valve and Sample Inlet Fitting Installation (8)
To install a rotary valve in the unheated compartment
Installing valves in the unheated compartment consists of the following steps:
• Installing the valve on the mounting bracket
• Installing the assemble into the unheated compartment under the main oven
• Installing the solenoid valve
• Connecting gas line
CAUTION Turn the actuator gas off before you start these procedures.
1. Install the valve on the mounting bracket
The unheated valve assembly will require the mounting bracket, included as part of the
assembly.
a. Choose the mounting holes in the bracket that will allow the actuator and the valve
to be oriented in a favorable direction.
b. Fix the valve to the bracket by using the two fixing screws. See