3
Components Installation
Installing Valves and Fittings into the Main Oven
48
TRACE 1310 Auxiliary Oven Instruction Manual
Figure 64. Main Oven Holes Layout (1)
b. Locate the desired valve position in the main oven and place the clamp ring over the
hole in the main oven, then secure the clamp ring with the two screws provided.
c. Loosen the dowel on the clamp ring, then insert the valve into the clamp ring.
d. Secure the valve tightening the dowel. See
Figure 65. Installing Diaphragm Valve
3. Connecting the gas line.
a. By using the proper nuts and ferrules, connect the gas line from the solenoid valve
manifold to the diaphragm valve. See
Valves Mounting Holes
Figure
65.
Figure
66.