Thermo Scientific TRACE 1300 Instruction Manual page 71

Auxiliary ovens
Hide thumbs Also See for TRACE 1300:
Table of Contents

Advertisement

Thermo Scientific
Figure 60. Actuator Installation (1)
b. Once the two square fittings on the actuator and valve are coupled, tighten the collar
on the valve. Be sure this fitting is very tight so the actuator will not rotate on the
valve standoff.
c. Slide the valve and actuator assembly up or down to the desired location and tighten
the upper collar to hold the valve and actuator in place.
IMPORTANT The installation of the Valve Actuator Cooling Kit (PN 19071023) is
recommended to preserve maximum actuators lifetime when TRACE 1310 Auxiliary
Oven is used at high temperature, between 150 °C and 270 °C.
For temperature above 270 °C High temperature actuators are recommended and the
cooling fans must not be used.
For further details refer to the
Installing Valves and Fittings into the Main Oven
Valve Actuator Cooling Kit Installation
TRACE 1310 Auxiliary Oven Instruction Manual
3
Components Installation
Cooling Fans
Note.
45