3
Components Installation
Installing the Secondary Oven Assembly
86
TRACE 1310 Auxiliary Oven Instruction Manual
Figure 123. Secondary Oven Installation (3)
4. Connect the ground wire on the bottom of the assembly using the proper fixing screw,
then connect the other end of the wire to the common ground point on the chassis of the
instrument. See
Figure
Figure 124. Secondary Oven Installation (4)
5. Remount and fix the protective grid. See
Figure 125. Secondary Oven Installation (5)
6. Connect the cables of the heater element and of the RTD Pt100 probe respectively to the
connectors marked J3 and J9 on the Aux Temp VO/Power Supply board.