Replacing the Column into the Secondary Oven
Thermo Scientific
This section provides instructions for replacing the column installed into the secondary oven.
To replace the column
1. Power off the GC and the TRACE 1310 Auxiliary Oven. See
TRACE 1310 Auxiliary Oven"
2. Remove the L-shape cover of the Auxiliary Oven.
3. Remove the L-shape cover of the main oven to access its interior.
4. Remove the protective grid from the back panel of the instrument by unscrewing the four
fixing screws. See
Figure
Figure 170. Column Replacement (1)
5. Disconnect the ends of column from the components at which the column is connected.
6. Remove the front panel of the secondary oven assembly. See
Figure 171. Column Replacement (2)
7. Carefully remove the column and its end from the secondary oven. See
on
page
121.
174.