b. Push the bracket against the studs, then secure the assembly tightening the studs.
3. Install the solenoid valves.
a. Install the solenoid manifold on the floor of the unheated compartment.
See the example of
Figure 81. Unheated Rotary Valve Installation (3)
b. Remove the rectangular shaped cap from the manifold using the Phillips screwdriver.
c. Secure the solenoid valve to the under plate using the two screws supplied with the
valve. See
Figure
82.
CAUTION Tighten screws evenly, taking care that the solenoid gasket rests on all surfaces
to prevent leaks.
Figure 82. Unheated Rotary Valve Installation (4)
d. Connect the cable from the solenoid valve to the proper timed event output jack of
the Events connector marked J24 on the VOBP-HRM board located over the power
section. See
Figure
Installing Valves and Fittings in the Unheated Compartment
Figure
81.
83.
TRACE 1310 Auxiliary Oven Instruction Manual
3
Components Installation
Solenoid Manifold
57