Meter Interfacing With Other Computer Programs - Thermo Scientific Orion Star A320 Series Manual

Electrochemistry portable meters
Table of Contents

Advertisement

Chapter 7
Data Transfer and Software Updates
|
88
Orion Star A320 Series Portable Meter User Manual
|

Meter Interfacing with Other Computer Programs

If using the meter USB port to connect with the computer, the Orion Star A200/A300 series
USB driver software (virtual COM port) must be installed on the computer and the meter should
be updated to software revision 3.04 or higher.
1. Power on the meter and connect the meter to the computer using the appropriate computer
interface cable.
2. Open a computer communication program that uses COM ports. Programs such as
Terminal and HyperTerminal apply.
3. Select the correct Virtual COM Port and connect. The following examples are from
HyperTerminal:
4. Verify that the meter export data setting is on. Match the baud rate set in the meter to the
computer setting. Selecting a higher baud rate will allow for faster data transfer.
5. Data will transfer from the meter to the computer, depending on the meter read type and
command settings.
Note: If the meter is powered off for any reason (auto-shut off, loss of power, etc.) the
connection will need to be reestablished by repeating the steps above.
Thermo Scientific

Advertisement

Table of Contents
loading

Table of Contents