4.2 K3HB-V Initial Setup Example (K3HB-VLC)
C Set the scaling value.
1. Set scaling input value A1 "inp. a 1" to "0. 000" and press the M [MODE]
Key.
2. Set scaling display value A1 "dsp. a 1" to "0" and press the M [MODE] Key.
3. Set scaling input value A2 "inp. a2" to "19. 999" and press the M [MODE]
Key.
4. Set scaling display value A2 "dsp. a2" to "1000" and press the M [MODE]
Key.
D Set the decimal point position.
1. Set the parameter "dp" to ",,. ,,," and press the M [MODE] Key.
E Set comparative set value H to 0.700 and set comparative set
value L to 0.500.
1. Return to the RUN level by pressing the L [LEVEL] Key for at least 1 s.
(Start operation.)
2. Press the M [MODE] Key repeatedly until the SV display status shows
.
H
3. Set the value to "0. 700" and press the M [MODE] Key. (The SV display
status will show
.)
L
4. Set the value to "0. 500" and press the M [MODE] Key. The setting
procedure is completed.
F Start actual operation.
1. Press the M [MODE] Key repeatedly to display the measurement values
and start actual operation.
Clearing Settings
If you become confused while setting the parameters and cannot
continue, all settings can be cleared so that you can start over.
Refer to "5.40 Initializing All Settings" (P.5-100) for information on
clearing all settings.
* Refer to "Section 5 Functions and Operations" for details on setting
parameters.