B Perform aging until the aging timer
reaches 0. (If the calibrator needs more
than 30 minutes of aging, extend the
aging until the conditions are met.)
• If the M [MODE] Key is pressed
while the aging timer is counting
down, the display skips to the
calibration
upper
display.
C Press the M [MODE] Key to display the
parameter for the calibration upper
limit.
• The parameter for the present input
type is displayed. Refer to the table
on P.6-9 for details on the relationship
between parameters and the input
type.
• The display will be as shown if "A" is
not included in the calculation, and
the calibration then will be for input
"B" (K3HB-S only).
D The calibrator applies a reference
signal that corresponds to the
calibration upper limit.
• Refer to the table on P.6-9 for
required reference signal values.
E Press the U [UP] Key.
• The reference signal is read and "T"
starts flashing.
F Press the U [UP] Key again to
temporarily register the calibration
upper limit.
G Repeat steps D to F to temporarily
register the calibration lower limit.
• When temporary registration has
been completed, the parameter for
registration "str" is displayed.
• Perform step L for the K3HB-H (TC
input).
H Press the S [SHIFT] Key to make the
SV display flash.
• The setting can be changed when the
SV display starts to flash.
6.2 User Calibration Operation
limit
parameter
M
U
U
S
adj
u
L
0
a4020
u
L
7f00
baa20
u
L
7f00
4000
u
L
7ef2
"T" is flashing.
4000
u
L
7ef2
"T" is lit.
str
u
L
no
str
u
L
no
6-7