Code
Description
16-EC1
The SA actuator of optical
block assembly Ch-2B can-
not detect the reference posi-
tion. (On side 1)
16-F41
The SA actuator of optical
block assembly Ch-3A can-
not detect the reference posi-
tion. (On side 1)
16-FC1
The SA actuator of optical
block assembly Ch-3B can-
not detect the reference posi-
tion. (On side 1)
ODS-D280U/ODS-D280F
Error remedy method
1.
Eject a cartridge.
2.
Startup the maintenance menu and perform
SA ACTUATOR.
•
When no error occurs, use the unit as
it is.
•
When an error occurs, go to step 3) in
Measures.
3.
Check the flexible flat cable connection be-
tween the sub base assembly and the DR-683
board.
If this problem still remains, the optical block as-
sembly (OP-2) or the DR-683 board may be de-
fective.
1.
Eject a cartridge.
2.
Startup the maintenance menu and perform
SA ACTUATOR.
•
When no error occurs, use the unit as
it is.
•
When an error occurs, go to step 3) in
Measures.
3.
Check the flexible flat cable connection be-
tween the sub base assembly and the DR-683
board.
If this problem still remains, the optical block as-
sembly (OP-3) or the DR-683 board may be de-
fective.
1.
Eject a cartridge.
2.
Startup the maintenance menu and perform
SA ACTUATOR.
•
When no error occurs, use the unit as
it is.
•
When an error occurs, go to step 3) in
Measures.
3.
Check the flexible flat cable connection be-
tween the sub base assembly and the DR-683
board.
If this problem still remains, the optical block as-
sembly (OP-3) or the DR-683 board may be de-
fective.
Refer to section
"4-29. Sub Base Assembly"
"6-4-8. SA ACTUATOR"
"4-29. Sub Base Assembly"
"6-4-8. SA ACTUATOR"
"4-29. Sub Base Assembly"
"6-4-8. SA ACTUATOR"
2-27