Sony ODS-D280U Service Manual page 58

Optical disc archive drive unit
Hide thumbs Also See for ODS-D280U:
Table of Contents

Advertisement

Main
Sub
Error
Code
Code
Logger
5A7
5B5
5B7
5C5
5C7
695
6A5
6B5
6C5
99F
AAF
BBF
CCF
92
595
597
5A5
5A7
5B5
5B7
ODS-D280U/ODS-D280F
Alarm
Alarm
Description
Status
Indica-
tor
Blinking
Cycle
0.25 sec.
Failure in Disc Drive LSI 0b (Base Drive) firmware upgrade was
detected during startup. (On side 0)
0.25 sec.
Communication between Disc Drive LSI 1a (Base Drive) and
BRIDGE (IC700 internal: BI-353 (ODS-D280U)/BF-92 (ODS-
D280F) board) cannot be established during startup. (On side 0)
0.25 sec.
Failure in Disc Drive LSI 1a (Base Drive) firmware upgrade was
detected during startup. (On side 0)
0.25 sec.
Communication between Disc Drive LSI 1b (Base Drive) and
BRIDGE (IC700 internal: BI-353 (ODS-D280U)/BF-92 (ODS-
D280F) board) cannot be established during startup. (On side 0)
0.25 sec.
Failure in Disc Drive LSI 1b (Base Drive) firmware upgrade was
detected during startup. (On side 0)
0.25 sec.
Communication between SYCPU (IC700 internal: BI-353
(ODS-D280U)/BF-92 (ODS-D280F) board) and 0a (Base
Drive) cannot be established during startup. Or, failure in 0a
(Base Drive) firmware upgrade was detected during startup. (On
side 0)
0.25 sec.
Communication between SYCPU (IC700 internal: BI-353
(ODS-D280U)/BF-92 (ODS-D280F) board) and 0b (Base
Drive) cannot be established during startup. Or, failure in 0b
(Base Drive) firmware upgrade was detected during startup. (On
side 0)
0.25 sec.
Communication between SYCPU (IC700 internal: BI-353
(ODS-D280U)/BF-92 (ODS-D280F) board) and 1a (Base
Drive) cannot be established during startup. Or, failure in 1a
(Base Drive) firmware upgrade was detected during startup. (On
side 0)
0.25 sec.
Communication between SYCPU (IC700 internal: BI-353
(ODS-D280U)/BF-92 (ODS-D280F) board) and 1b (Base
Drive) cannot be established during startup. Or, failure in 1a
(Base Drive) firmware upgrade was detected during startup. (On
side 0)
0.25 sec.
A firmware error of Disc Drive LSI 0a (Base Drive) detected an
abnormality. (On side 0)
0.25 sec.
A firmware error of Disc Drive LSI 0b (Base Drive) detected an
abnormality. (On side 0)
0.25 sec.
A firmware error of Disc Drive LSI 1a (Base Drive) detected an
abnormality. (On side 0)
0.25 sec.
A firmware error of Disc Drive LSI 1b (Base Drive) detected an
abnormality. (On side 0)
0.25 sec.
Communication between Disc Drive LSI 2a (Base Drive) and
BRIDGE (IC700 internal: BI-353 (ODS-D280U)/BF-92 (ODS-
D280F) board) cannot be established during startup. (On side 1)
0.25 sec.
Failure in Disc Drive LSI 2a (Base Drive) firmware upgrade was
detected during startup. (On side 1)
0.25 sec.
Communication between Disc Drive LSI 2b (Base Drive) and
BRIDGE (IC700 internal: BI-353 (ODS-D280U)/BF-92 (ODS-
D280F) board) cannot be established during startup. (On side 1)
0.25 sec.
Failure in Disc Drive LSI 2b (Base Drive) firmware upgrade was
detected during startup. (On side 1)
0.25 sec.
Communication between Disc Drive LSI 3a (Base Drive) and
BRIDGE (IC700 internal: BI-353 (ODS-D280U)/BF-92 (ODS-
D280F) board) cannot be established during startup. (On side 1)
0.25 sec.
Failure in Disc Drive LSI 3a (Base Drive) firmware upgrade was
detected during startup. (On side 1)
Continued