SECTION 8: SPECIFICATIONS

8.1

PERFORMANCE

8.1
Performance
8.2
Electrical
8.3
Environmental
8.4
Physical
Measurement Range
SpO
2
0–100%
Pulse Rate
20–250 beats per minute (bpm)
Accuracy
SpO
2
Adults
± 2 digits
70–100%
0–69%
unspecified
Neonates
± 2 digits
70–100%
0–69%
unspecified
Note:
Accuracies are expressed as plus or minus "X" digits (oxygen
saturation percentage points) between saturations of 70-100%. This
variation equals plus or minus one standard deviation (1SD), which
encompasses 68% of the population. All accuracy specifications are
based on testing the subject monitor on healthy adult volunteers in
induced hypoxia studies across the specified range. Adult accuracy is
determined with Oxisensor® II D-25 sensors. Neonatal accuracy is
determined with Oxisensor II N-25 sensors. In addition, the neonatal
accuracy specification is adjusted to take into account the theoretical
effect of fetal hemoglobin in neonatal blood on oximetry
measurements.
Pulse Rate
± 3 bpm
20–250 bpm
Note:
Accuracy is expressed as plus or minus 3 bpm across the display
range. This variation equals plus or minus one standard deviation
(1SD), which encompasses 68% of the population.