S p e c i f i c a t i o n s

Performance

N-595
SpO
2
Pulse Rate
Perfusion Range
Accuracy and Motion Tolerance
Saturation
1
Without Motion - Adult
Without Motion - Neonate
With Motion - Adult and Neonate
3
Low Perfusion
1
Adult specifications are shown for O
the N-595. Neonate specifications are shown for O
N-595. Saturation accuracy will vary by sensor type. Refer to the Sensor Accuracy
Grid.
2
Applicability: O
M
XI
AX
sensors.
3
Specification applies to monitor performance and was validated with Biotek and
Nellcor simulators.
Pulse Rate
3
Without Motion
With Motion
3
Low Perfusion
3
Specification applies to monitor performance and was validated with Biotek and
Nellcor simulators.
Measurement Range
1% to 100%
20 beats per minute (bpm) to 250 bpm
0.03% to 20%
70 to 100% ±2 digits
1
70 to 100% ±3 digits
2
70 to 100% ±3 digits
70 to 100% ±2 digits
M
MAX-A and MAX-N sensors with
XI
AX
MAX-A, MAX-AL, MAX-P, MAX-I, and MAX-N
20 to 250 ± 3 digits
normal physiologic range (e.g.,
55 - 125 bpm) ± 5 digits
20 to 250 ± 3 digits
M
MAX-N sensors with the
XI
AX
75