Routine Maintenance; Cleaning; Periodic Safety Checks; Functional Checks - Nellcor OxiMax N-595 Service Manual

Hide thumbs Also See for OxiMax N-595:
Table of Contents

Advertisement

R o u t i n e M a i n t e n a n c e

Cleaning

Periodic Safety Checks

Functional Checks

N-595
WARNING: Do not spray, pour, or spill any liquid on the N-595, its
accessories, connectors, switches, or openings in the chassis.
For surface-cleaning and disinfecting follow your institution's procedures or:
The N-595 may be surface-cleaned by using a soft cloth dampened with
either a commercial, nonabrasive cleaner or a solution of 70% alcohol in
water, and lightly wiping the surfaces of the monitor.
The N-595 may be disinfected using a soft cloth saturated with a 10%
solution of chlorine bleach in tap water.
Before attempting to clean an SpO
enclosed with the O
-M
XI
specific to that sensor.
The N-595 requires no calibration.
The battery should be replaced at least every 2 years. See Battery Replacement on
page 67.
The following checks should be performed at least every 24 months by a qualified
service technician.
1. Inspect the equipment for mechanical and functional damage.
2. Inspect safety labels for legibility. If the labels are damaged, contact Nellcor's
Technical Services Department, 1.800.635.5267, or your local Nellcor
representative.
If the monitor has been visibly damaged or subjected to mechanical shock (for
example, if dropped), immediately perform the performance tests. See
Performance Tests on page 11.
O
-M
sensor, read the directions for use
2
XI
AX
sensor. Each sensor model has cleaning instructions
AX
9