R o u t i n e M a i n t e n a n c e

Cleaning

Periodic Safety Checks

Functional Checks

O
M
N-65 Service Manual
XI
AX
WARNING: Do not spray, pour, or spill any liquid on the O
accessories, connectors, switches, or openings in the enclosure as this may
damage the monitor.
For surface-cleaning and disinfecting follow your institution's procedures or:
The O
M
N-65 may be surface-cleaned by using a soft cloth
XI
AX
dampened with either a commercial, nonabrasive cleaner or a solution of
70 percent alcohol in water, and lightly wiping the surfaces of the
monitor.
The monitor may be disinfected using a soft cloth saturated with a ten
percent solution of chlorine bleach in tap water.
Before cleaning an SpO
with the O
M
sensor. Each sensor model has cleaning instructions specific to
XI
AX
that sensor.
The O
M
N-65 does not require calibration.
XI
AX
A qualified service technician should perform the following safety checks every
24 months.
To perform periodic safety checks:
1. Inspect the equipment for mechanical and functional damage.
2. Inspect safety labels for legibility. If the labels are damaged, contact Nellcor's
Technical Services Department, 1.800.635.5267, or your local Nellcor
representative.
If the monitor has been visibly damaged or subjected to mechanical shock (for
example, if dropped), immediately perform the performance tests. See
Performance Tests on page 17.
Every two years a qualified service technician should perform the electrical safety
tests detailed in Safety Tests on page 37. If the unit fails these electrical safety
tests, refer to Troubleshooting on page 39.