Section 2: Routine Maintenance

2.1 Cleaning

2.2 Periodic Safety and Functional Checks

2.3 Functional Checks
2.4 Batteries
2.5 Environmental Protection
WARNING: Do not spray or pour any liquid on the monitor or its accessories. Do
not immerse the N5500 or its accessories in liquid or clean with caustic or
abrasive cleaners.
2.1 Cleaning
To clean the N5500, dampen a cloth with a commercial, nonabrasive cleaner and wipe
the exterior surfaces lightly.
CAUTION: Do not allow any liquids to come in contact with the power connector
or switches or to penetrate connectors or openings in the instrument.
Note: For cables, sensors and cuffs, follow the cleaning instructions in the directions
for use that accompany these accessories.
Note: If liquid is spilled on the monitor, clean and dry thoroughly before reuse.
Note: If in doubt about monitor safety, refer the unit to qualified service personnel.
For surface-cleaning, follow your institution's procedures or:
l
2.2 Periodic Safety and Functional Checks
The N5500 requires no routine service or calibration other than cleaning and battery
maintenance. The following performance verification tests may be used following
repair or during routine maintenance (if required by your local institution).
1. Inspect the exterior of the N5500 for damage.
2. Inspect labels for legibility. If the labels are not legible, contact Nellcor Technical
Services Department or your local Nellcor representative.
3. Verify that the unit performs properly as described in Performance Verification
section.
The N5500 may be surface-cleaned by using a soft cloth dampened with
either a commercial, nonabrasive cleaner or a solution of 70% alcohol in
water, and lightly wiping the top, bottom, and front surfaces of the monitor
lightly.
N5500 Service Manual
7