Shutdown

Error ID: 0063
An external instrument has generated a shutdown signal on the remote line.
The module continually monitors the remote input connectors for status
signals. A LOW signal input on pin 4 of the remote connector generates the
error message.
Probable cause
1
2
3
4
1260 Infinity QuatPump VL User Manual
Leak detected in another module with a CAN
connection to the system.
Leak detected in an external instrument with
a remote connection to the system.
Shut-down in an external instrument with a
remote connection to the system.
The degasser failed to generate sufficient
vacuum for solvent degassing.
Error Information
General Error Messages
Suggested actions
Fix the leak in the external instrument before
restarting the module.
Fix the leak in the external instrument before
restarting the module.
Check external instruments for a shut-down
condition.
Check the vacuum degasser for an error
condition. Refer to the Service Manual for the
degasser or the 1260 pump that has the
degasser built-in.