Setting Up and Running an Acquisition Method

Setting Up and Running an Acquisition Method

1 Set up the method:
a Type the values and settings for each of the tabs below.
b Optional. To download the settings to the instrument, click Apply.
c To save the method, click Save As.
d Name the method and click OK.
2 Enter values for all of the LC modules configured for the instrument.
Make sure when you type the MS parameters on the next page that the tune file
NOTE
is the one that you want to use with the acquisition.
3 Enter the TQ parameter values.
a Select the Scan Type from the list in the Time segments table. The Scan
Agilent 6475 Triple Quad LC/MS System User Guide
segments table is cleared when you change the Scan Type. The
parameters available on the right change depending on the Scan Type for
the selected Time segment.
If you are changing the Scan Type from MRM to Dynamic MRM or to
Triggered MRM, you can copy and paste the transitions from the original
Scan segments table to the Clipboard and then to the new Scan segments
table. See the online Help for more information.