Initial Setup - Teledyne LabMaster 10 Zi Series Operator's Manual

Hide thumbs Also See for LabMaster 10 Zi Series:
Table of Contents

Advertisement

LabMaster 10 Zi Oscilloscopes
All True
Any True
All Q1 to Q4 Or All Q5 to Q8
P
/F
ASS
AIL

Initial Setup

1.
Touch Analysis → Pass/Fail Setup... on the menu bar.
2.
Touch the Actions tab.
3.
Touch the Enable Actions checkbox. This causes selected actions
to occur on your waveform's passing or failing a test.
4.
Touch the Summary View checkbox to enable a line of text (as
follows) showing a concise status of the last waveform and
keeping a running count of how many sweeps have passed.
5.
Touch inside the Pass If control, and select a Boolean condition
from the choices shown.
6.
If you want to set up the test to end after a finite number of
sweeps, touch the Stop Test checkbox. Then touch inside the
After data entry control and provide a value using your preferred
input control method. Additional information on using the touch
screen controls can be found in the Dialog Area.
7.
Use the
the dialog to set actions taking place when your waveform passes
or fails the test.
8.
The Then section of the dialog provides stop test, sound alarm,
print result, emit pulse, or save (the waveform) actions. If you
156
Pass/Fail Boolean Conditions
T
S
ESTING
ETUP
Pass or
All False
Any False
Any Q1 to Q4 And Any Q5 to Q8
Fail button on the If section of
LM10Zi-GSM-E Rev A