Viewing A Track - Teledyne LabMaster 10 Zi Series Operator's Manual

Hide thumbs Also See for LabMaster 10 Zi Series:
Table of Contents

Advertisement

LabMaster 10 Zi Oscilloscopes
the lower-right side of the screen).
NOTE: Turning off a trace for which trend data is being collected
resets the trend. If it is necessary to continue data collection for
the trend, first create a zoom trace of the channel trace (before
turning off the channel trace).

Viewing a Track

This procedure explains how to view a Track of a math function or
measurement parameter applied to a waveform. A Track shows... See
Track vs. Trend (on page 142) for a better understanding of the
information a Track provides.
1.
In the menu bar, touch Measure → Measure Setup....
2.
Touch a parameter tab Px and its corresponding dialog is shown.
3.
Now, on the dialog, touch inside the Source1 data entry field and
select an input waveform from the pop-up menu.
4.
Touch inside the Measure data entry field and select a parameter
from the pop-up menu.
5.
Touch the Track button at the bottom of the dialog and select a
math function location (Fx) on which to store the Track display.
148
LM10Zi-GSM-E Rev A

Advertisement

Table of Contents
loading