Teledyne API Model 401 O

4 SETUP MODE

This section describes the setup variables which are used to configure the analyzer.

4.1 Setup Mode Operation

All the setup variables are stored in the analyzer's EEPROM and are retained during power off
and even when new software revisions are installed.
If a variable is modified, but ENTR is not pressed, the variable will
not be changed and the analyzer will beep when EXIT is pressed.

4.2 Setting Automatic Multipoint (ASEQ) Check

Automatic sequence of multipoint calibration can be programmed to select one zero and five
span points 20%, 40%, 60%, 80%, and 100% of selected (see Section 3.4 How to set
concentration) O
Likewise, select AZS for zero and one span point (selected O
zero and two span points 25% and 100% of selected O
pre-programmed automatic sequence. If ASEQ is selected, then front panel's yellow status LED
will be turned on.

4.3 Setting Automatic Zero/Span Duration

Duration of each zero/span calibration step can be programmed by selecting SETUP-EVNT-
WAIT. Enter the duration between 1 - 20 minutes for the zero check and then press ENTR.
similarly enter the duration for span check and press ENTR.
Teledyne API recommends that the durations are set at least 10
minutes to allow for an accurate and reliable ZERO or SPAN value
Automatic zero/span calibration will be initiated once per day on a timed basis. Set starting time
of the automatic calibration by selecting SETUP-EVNT-TIME. Time should be between 00:00
- 23:59.
Starting time of the automatic cal. can be shifted ± 60 min to avoid missing data of remote
analyzer during same time of the day. Set desired shift time by selecting SETUP-ENVT-SHFT.
Photometric Calibrator Instruction Manual, 01124, Rev. J1
3
NOTE
concentration by pressing sequence of SETUP-EVNT-ASEQ-AZ5S.
3
NOTE
to be established.
concentration), select AS2S for
3
concentration, or press OFF to disable
3
4-1