LabMaster 10 Zi Oscilloscopes
2.
Place a checkmark next to the module(s) you want to calibrate
and touch the ChannelSync Cal button. The ChannelSync
Calibration Wizard opens.
NOTE: Use the Ref Channel control to select the specific channel
you wish to use as a reference when measuring against other
channels during your calibration.
3.
Select the inputs you want to measure (for example, Input A or
Input B). You can select Active Input to make measurements on
the currently selected active input of each channel. Touch the
Next button.
4.
Connect the Fast Edge calibration signal (located on the front of
the Acquisition Module) to the Input you provided as the
reference input, and then touch the Next button. The
ChannelSync™ Calibration measurements are performed on the
input you entered (in this case, C1).
5.
Connect your Fast Edge calibration signal to the next input (in this
case, C2), and then touch the Next button. Continue this process
for each channel. As each channel is calibrated, a respective trace
descriptor is shown for the channel under the grid display area.
6.
When you have finished all channels, a message indicates all
measurements have been performed. Touch the Finish button.