LabMaster 10 Zi Oscilloscopes
Measurement Mode – Searches from 2 to 64 bits; ideal for
bursted patterns where a PLL cannot lock.
Bus Pattern Mode – Searches from 2 to 64 bits; enhances MSO
search capabilities.
Parameter Measurements
Besides parameter measurements made during acquisition, post-
acquisition measurements can also be made. The number of parameters
available depends on the options loaded on your instrument.
Measurements are made only on the events defined by your filter
(search criteria). A Filter Wizard is provided to quickly establish
statistical criteria such as ±1, 3, or 5 sigma.
Sampling Mode
Whenever WaveScan is enabled, the instrument reverts to Real-time
sampling mode.

Customization Overview

You can customize the instrument to your needs by using the power of
programs such as Excel™, Mathcad™, and MATLAB™ or by scripting in
VBS. Add your own parameters, functions, display algorithms, or other
routines to the oscilloscope user interface and make them an integral
part of the instrument's application environment. Whichever method
you use, the results appear on the instrument's display together with
the acquired signals. This offers a tremendous ability to solve unique
problems for a large range of applications, with comparatively little
effort from you.
Instrument customization provides these important capabilities:
You can export data to external programs and display the results
on the instrument, without leaving the oscilloscope environment.
Once the result is returned, you can perform additional
oscilloscope operations—such as measuring with cursors,
164