LabMaster 10 Zi Oscilloscopes
Turn this knob to adjust the level of the trigger source or the highlighted
trace. Level defines the source voltage at which the trigger will generate
an event: a change in the input signal that satisfies the trigger
conditions.
Alternatively, in the Trigger dialog, you can touch inside the Level
control and provide a value using your preferred input control method.
Additional information on using the touch screen controls can be found
in the Dialog Area. Quickly set a level of zero volts by touching the Zero
Level button.
An arrow on the right side of the grid shows the threshold position. This
arrow is only visible if the trigger source is displayed.

Trigger Setup

Based on your Trigger Type selection, sections of the main Trigger
dialog and additional dialogs altogether vary.
Most trigger types have Type, Setup, and Level selections. The following
sections explain some of the standard setup configurations for different
trigger types.
1.
Make your Type selection by touching the button corresponding
to the desired trigger.
2.
Touch inside the trigger Source control for your first Setup
configuration and select a source on which to trigger.
3.
Touch inside the Coupling control and select a coupling mode.
Coupling refers to the type of signal coupling at the input of the
trigger circuit.