Signal Views

WaveScan provides distinct views of your signal:
Source View highlights all occurrences of edges that meet your
criteria.
Scan Overlay places all captured edges one on top of the other
in a separate grid. You can apply monochromatic persistence in
this view.
Scan Histogram provides a statistical view of edges that meet
your search criteria.
Zoom View allows you to expand a waveform feature vertically
and horizontally; to apply further processing (using Measure); to
Store it, or to apply a descriptive Label to the feature.
Note: The number of grids displayed varies from one to three grids
depending on which views are enabled. WaveScan handles this function
automatically, and there is no option to move traces from one grid to
another, as would be the case under normal (non-WaveScan) operation.

Search Modes

Use the Mode control to select an appropriate Search to locate
anomalies during acquisition. When each of the following modes are
selected, a corresponding right-hand dialog provides additional controls.
The following modes are available:
Edge Mode – Designed for detecting the occurrence of edges;
selectable slope and level.
Non-monotonic Mode – For detecting threshold re-crosses;
selectable slope, hysteresis, and level.
Runt Mode – For detecting pulses failing to cross a threshold;
selectable polarity and thresholds.
Measurement Mode – For filtering and performing parameter
measurements.