LabMaster 10 Zi Oscilloscopes

TriggerScan

TriggerScan is a debugging tool (available for any trigger type) that helps
you quickly find rare waveform glitches and anomalies. With
TriggerScan, you can build a list of trigger setups to look for rare events
and automatically sequence through each one. TriggerScan can use any
type of trigger setup available including edge, width, and qualify as well
as Smart Triggers (such as, glitch and runt triggers). TriggerScan
automates two key processes in triggering rare events:
Trains the system by looking at normal acquired waveforms.
During the training, the oscilloscope analyzes the waveforms to
determine what waveforms normally look like. Using this
information, it generates a list of smart trigger setups to trigger
on abnormal situations.
Loads the smart trigger setups from the Trainer and cycles
through these. As triggers occur, they are overlaid on the screen.
All acquisition settings are preserved and you can use all the
functions of the oscilloscope to find the root cause of these
anomalies.