In Range and Out of Range conditions can be further specified by
choosing to set:
Limits and assigning Upper and Lower Values to the range.
OR
Delta and assigning Nominal width and Delta.
NOTE: Width Condition settings are summarized on the far right of the
dialog.

Holdoff by Time or Events

Holdoff is an additional condition of Edge and Pattern triggers. It can be
expressed either as a period of time or an event count. Holdoff disables
the trigger circuit for a given period of time or number of events after
the last trigger occurred. Events are the number of occasions on which
the trigger condition is met. The trigger resumes when the holdoff has
elapsed and its other conditions are met.
Select the holdoff by touching the Time or Events buttons on the
Holdoff dialog.
Now, click the corresponding control and provide a value.
Use holdoff to obtain a stable trigger for repetitive, composite
waveforms. For example, if the number or duration of sub-signals is
known you can disable them by choosing an appropriate holdoff value.
Qualified triggers operate using conditions similar to holdoff.