Recorder Output

The rear panel Recorder Output connectors (1 and 2) produce a dc
voltage that corresponds to the power level in Watts of the channel
depending on the measurement mode. This dc voltage ranges from 0 to
+1 Vdc. The output impedance is typically 1 kΩ. Channel and display
offsets, and duty cycle have no effect on the Recorder Outputs.
For example, the Recorder Outputs can be used to:
• Record swept measurements
• Level an output from a source using external leveling
• Monitor the output power
To access the Recorder menu press
allows you to switch the Recorder Output signal either on or off. The
Max Power
represent the 1 Vdc maximum and 0 Vdc minimum output voltage of the

Recorder Output.

Procedure
Set the recorder output as follows:
Ensure you have selected the channel you want to set up.
N O T E
1 Press
2 Use the
3 Press
N1911A/1912A P-Series Power Meters User's Guide
Min Power
and
softkeys allow you to scale the power levels to
Meas Select
,
keys to highlight the Rec o/p: setting field.
and
to reveal the Recorder option menu.
General Power Meter Functions
, and enable Rec o/p. This
to display the Measurement Setup menus.
2
99