8480 Series Sensors Connection Requirements

Table 7-12 8480 Series connection requirements

N1911A/1912A P-Series Power Meters User's Guide
Sensor
8481A
These power sensors connect directly to the POWER REF.
8481H
8482A
8482H
8481D
Prior to calibration, an Agilent 11708A 30 dB reference attenuator
8484A
should be connected between the power sensor and the POWER
REF. Remove this attenuator from the power sensor input before
making measurements.
This power sensor requires a 75 Ω (f) to 50 Ω (m) N-Type adapter
8483A
(1250-0597) to connect to the POWER REF. Remove this adapter
before making measurements.
R8486A
These waveguide power sensors have two connectors. Use the
Q8486A
N-Type connector to calibrate the power meter.
V8486A
W8486A
R8486D
Q8486D
8481B
These power sensors are configured with an attenuator. Prior to
8482B
calibration this attenuator must be removed. Replace the attenuator
before making measurements.
This power sensor requires an APC 3.5 (f) to 50 Ω (m) N-Type
8485A
adapter (08485-60005) to connect to the POWER REF.Remove this
adapter before making measurements.
8485D
Prior to calibration, an Agilent 11708A 30 dB reference attenuator
and an APC 3.5 (f) to 50 Ω (m) N-Type adapter (08485-60005) should
be connected between the power sensor and the POWER REF.
Remove this attenuator and adaptor before making measurements.
Using 8480 Series Power Sensors
Connection Requirements
7
177