11 Use the numeric keys to enter the power level you want to generate a 0
12 Press
The recorder output can be disabled and re-enabled by checking the Rec o/p: setting field.
N O T E
The highest power you are going to measure is used to determine the value which you
N O T E
should set for the Recorder Output maximum setting. For example, if you are measuring a
power less than 1 mW and greater than 100 μW, then set the recorder maximum value to
1 mW.

Table 2-8 Ranges of recorder output setting

Log
Lin
Log
Lin
N1911A/1912A P-Series Power Meters User's Guide
Vdc output in the Recorder Minimum pop- up and press
key to close the Measurement Setup screen.
50
40
100 W
10 W
–10
–20
100 mW
10 mW
General Power Meter Functions
30
20
1 W
100 mW
–30
–40
1 mW
100 nW
dBm
.
10
0
10 mW
1 mW
–50
–60
10 nW
1 nW
2
101