7
Using 8480 Series Power Sensors
Editing frequency dependent offset tables requires the following steps:
1 Identify and select the table you want to edit
2 Rename the table
3 Enter the frequency and offset pairs
4 Save the table
Procedure
First select the table you want to edit as follows:
1 Press

Figure 7-131"Sensor Tbls" screen

2 Choose the table you want to edit using the
188
Tables
,
screen.
Edit Table
to display the Edit Cal screen as shown in
Sensor Cal Tables
,
N1911A/1912A P-Series Power Meters User's Guide
to display the Sensor Tbls
and
keys. Press
Figure 7-
132.