System Specifications And Characteristics - Agilent Technologies N1911A User Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

System Specifications and Characteristics

The video bandwidth in the meter can be set to High, Medium, Low and
Off. The video bandwidths stated in the table below are not the 3 dB
bandwidths, as the video bandwidths are corrected for optimal flatness
(except the Off filter). Refer to
flatness response of the High, Medium, Low and Off filters. The Off video
bandwidth setting provides the warranted rise time and fall time
specification and is the recommended setting for minimizing overshoot on
pulse signals.
Rise Time, Fall Time, and Overshoot versus Video Bandwidth
Settings
Parameter
Rise Time/ Fall Time
10
Overshoot
For option 107 (10 m cable), add 5 ns to the rise time and fall time specifications.
N O T E
N1911A/1912A P-Series Power Meters User's Guide
Low:
5 MHz
9
<
56 ns
-
9
Specified as 10 % to 90 % for rise time and 90 % to 10 % for fall time on a 0 dBm pulse.
10
Specified as the overshoot relative to the settled pulse top power.
Specifications and Characteristics
Figure 11- 159
Video Bandwidth Setting
Medium:
High:
15 MHz
30 MHz
<
25 ns
13 ns
-
-
for information on the
Off
< 500 MHz
> 500 MHz
<
36 ns
<
<
5 %
11
13 ns
5 %
267

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

1912aN1912a

Table of Contents