Figure 11-158Measured Rise Time Percentage Error Versus Signal Under Test Rise Time - Agilent Technologies N1911A User Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

% Error
35
30
25
20
15
10
5
0

Figure 11-158Measured rise time percentage error versus signal under test rise time

Although the rise time specification is ≤13 ns, this does not mean that the
P- Series meter and sensor combination can accurately measure a signal
with a known rise time of 13 ns. The measured rise time is the root sum
of the squares (RSS) of the signal under test rise time and the system rise
time (13 ns).
N1911A/1912A P-Series Power Meters User's Guide
15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100
(
(
SignalUnderTestRiseTime
(
MeasuredRiseTime SignalUnderTestRiseTime
-------------------------------------------------------------------------------------------------------------------------------------- -
(
SignalUnderTestRiseTime
Specifications and Characteristics
Signal under test rise time (ns)
2
)
(
+
SystemRiseTime
×
)
2
)
)
)
100
11
261

Hide quick links: