Using N8480 Series Power Sensors
8
2
The Agilent 8482B and Agilent 8482H power sensors use the same data as the Agilent 8482A.
There are also ten sensor calibration tables named CUSTOM_0 through
CUSTOM_9. These tables do not contain any data when the power meter
is shipped from the factory.
Editing frequency dependent offset tables requires the following steps:
1 Identify and select the table you want to edit
2 Rename the table
3 Enter the frequency and offset pairs
4 Save the table
Procedure
First select the table you want to edit as follows:
Tables
Sensor Cal Tables
1 Press
,
,
to display the Sensor Tbls
screen.

Figure 8-144"Sensor Tbls" screen

2 Choose the table you want to edit using the
and
keys. Press
Edit Table
to display the Edit Cal screen as shown in
Figure 8-
145.
N1911A/1912A P-Series Power Meters User's Guide
213