8
Using N8480 Series Power Sensors
You can reduce the steps required to carry out the zero and calibration procedure as
N O T E
follows:
1 Connect the power sensor to the POWER REF output.
2 Press
Now set the frequency of the signal you want to measure. The power
meter automatically selects the appropriate calibration factor.
6 Press
7 Use the
Figure 8-137Frequency pop-up
8 Confirm your choice by pressing
9 Press
Now make the measurement.
10 Connect the power sensor to the signal to be measured.
The corrected measurement result is displayed.
202
Zero + Cal
and
Zero + Cal
Zero + Cal A
,
. On dual channel meters select the required channel.
and
to display the Frequency pop- up. Use the numeric keypad to
enter the required value in the Frequency pop- up window.
key to close the Channel Setup screen.
. (For dual channel meters, press
Zero + Cal B
or
as required).
keys to highlight the Frequency value field and press
MHz
or
N1911A/1912A P-Series Power Meters User's Guide
GHz
.