Figure 2-38 Gate Setup Screen - Agilent Technologies N1911A User Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

2
General Power Meter Functions
Figure 2-37
N O T E
Procedure
This procedure uses the Gates Setup under the Channel Setup menu. Alternatively, you
N O T E
can use a more visual method, when in the Graphical Trace Mode (Gate Control Menu), to
set the channel's Gate Control. (See
• Press

Figure 2-38 Gate Setup screen

1 Highlight the Gate Start you want to configure using the
The gate start time is relative to the trigger event. Positive values set a measurement gate,
N O T E
to a maximum time of 1 second, after the trigger. Negative time gate start values set a
measurement gate, to a maximum time of 1 second, before the trigger.
58
shows measurement results using this example of the gates placement.
Gates Setup
. The Channel Gate Setup screen is displayed.
keys.
"Setting the Trace Display"
N1911A/1912A P-Series Power Meters User's Guide
on page 118).
,
,
and