8
Using N8480 Series Power Sensors
Introduction
The N8480 Series power sensors is a replacement for the 8480 Series
power sensors (excluding the D- model sensors) with build- in Electrically
Erasable Programmable Read- Only Memory
The N8480 Series power sensors is used for measuring the average power
supplied by RF or microwave source or a device- under- test (DUT). The
N8480 Series power sensors places a 50 ohm load on the RF or microwave
source. The power meter indicates the power dissipated in this load in W
or dBm.
The N8480 sensors (excluding Option CFT) measure power levels from
–35 dBm to +20 dBm (316 nW to 100 mW), at frequencies from 10 MHz to
33 GHz and have two independent power measurement path (upper and
lower range).
Table 8-14 Power range in the Range setting
excluding Option CFT
N8481/2B excluding
N8481/2H excluding
Meanwhile, the N8480 sensors with Option CFT only measure power levels
from –35 dBm to +20 dBm (1 µW to 100 mW) in single range.
Similiar to the E- Series power sensors, the N8480 Series power sensors
are also equipped with EEPROM to store sensor's characteristics such as
model number, serial number, linearity, temperature compensation,
calibration factor
stored in EEPROM is not applicable for N8480 Series power sensors with
196
Sensor
AUTO (Default)
N8481/2/5/7/8A
and N8486AQ/AR
AUTO (Default)
Option CFT
AUTO (Default)
Option CFT
1
and so forth. However, the calibration factor table
Range Setting
–35 dBm to –1 dBm
LOWER
–35 dBm to –1 dBm
2
UPPER
–5 dBm to +29 dBm
LOWER
–5 dBm to +29 dBm
2
UPPER
–15 dBm to +17 dBm
LOWER
–15 dBm to +17 dBm
2
UPPER
N1911A/1912A P-Series Power Meters User's Guide
(
1
EEPROM)
.
Lower Range
–1 dBm to +20 dBm
-
–35 dBm to +20 dBm
+29 dBm to +44 dBm
-
+29 dBm to +44 dBm
+17 dBm to +35 dBm
-
+17 dBm to +35 dBm
Upper Range
-
-
-