8
Using N8480 Series Power Sensors
Editing/Generating Sensor Calibration Tables
Predefined sensor calbration factor table stored in power meter is not applicable for
N O T E
Agilent N8480 Series power sensors with Option CFT. Therefore, users are required to
create a new sensor calibration table for the sensors when a sensor calibration table is
needed.
To help achieve the best accuracy in your measurement you can enter the
values supplied for the sensors you are using by editing the installed
sensor calibration tables or by generating your own custom tables.
You cannot delete any of the 20 existing calibration tables or create any
additional tables. However, you can edit or delete the content of each
table. If you need another table you should edit and re- name one of the
tables.Each calibration table can contain a maximum of 80 frequency
points.
To view the calibration tables currently stored in the power meter, press
shown in

Table 8-16 Installed power sensor models

212
,
,
Tables
Sensor Cal Tables
Figure 8-
141.
Table
Sensor Model
0
DEFAULT
1
8481A
2
8482A
3
8483A
4
8481D
1
DEFAULT is a sensor calibration table where the reference calibration factor and calibration
factors are 100%. This sensor calibration table can be used during the performance testing of the
power meter.
. The Sensor Tbls screen is displayed as
Table
1
5
6
2
7
8
9
N1911A/1912A P-Series Power Meters User's Guide
Sensor Model
8485A
R8486A
Q8486A
R8486D
8487A