Setting The Cable Short/Long - Agilent Technologies N1911A User Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

Setting the Cable Short/Long

When connecting the power sensor with any Agilent power sensor cable
with length of 10 m (32 ft) and above, you are required to change the
cable setting from Short (default setting) to Long, or an error will occur
when the power sensor is connected.
Short
This is the default setting and for sensor cable length less than or equal to
10 m (32 ft).
Long
This setting is for sensor cable length more than 10 m (32 ft). The
maximum supported sensor cable length is 31 m (100 ft).
To access the cable setting,
1 Press
2 In the Service menu, press
The LCB indicator will appear in blue at the bottom of the display to
indicate that the long cable option is selected as shown in
Figure 2-91 Short/long cable option
N1911A/1912A P-Series Power Meters User's Guide
1 of 2
,
between the short or long cable options.
General Power Meter Functions
Service
and select
.
1 of 2
and select
Cable Short/Long
to toggle
Figure 2-
2
91.
113

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

1912aN1912a

Table of Contents