Power Meter Configuration

The P- Series power meters automatically recognize an E- Series E9300
power sensor when it is connected. The sensor calibration data is
automatically read by the power meter. The power meter also configures
the auto- averaging settings shown in
sensor characteristics.

Figure 5-113E9300 E-Series auto-averaging settings

These values are valid only for the power meter channel connected with E-Series E9300
N O T E
power sensor. You can also configure the settings manually – refer to
Results with TDMA Signals"
N1911A/1912A P-Series Power Meters User's Guide
E9300/1/4A
E9300/1H
10 dBm
20 dBm
2 dBm
12 dBm
-4 dBm
6 dBm
-10 dBm
0 dBm
-20 dBm
-10 dBm
-30 dBm
-20 dBm
-40 dBm
-30 dBm
-50 dBm
-40 dBm
on page 160 if required.

Using E9300 E-Series Power Sensors

Figure 5- 113
Maximum
E9300/1B
Sensor Power
40 dBm
32 dBm
26 dBm
20 dBm
10 dBm
0 dBm
-10 dBm
-20 dBm
Minimum
Sensor Power
to suit the power
Resolution Setting
1
2
3
4
1
1
1
4
1
1
4
16
1
1
8
32
1
4
16
128
1
16
64
128
1
1
1
4
1
1
2
16
1
2
16
64
4
16
128
256
32
64
256
256
"Achieving Stable
5
153

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

1912aN1912a

Table of Contents