2
General Power Meter Functions
• Press
The Zeroing pop- up is displayed when a zero occurs.
When you are performing measurements, this may cause delays in obtaining results.
N O T E
During this time, it is recommended to disable the automatic zeroing.
Manual Zeroing
To manually zero the power meter and sensor:
• Press
On dual channel meters you can zero each channel independently or both channels
N O T E
sequentially by pressing the
Calibration
Calibration sets the gain of each power meter channel and P- Series sensor
combination. This is achieved without the need to connect it to the
1.0 mW Power Reference. The power meter can be set to automatically
calibrate on- the- fly while you are using it. Alternatively, you can perform
the calibration manually. The reference calibration factor is automatically
set for all P- Series power sensors.
During calibration, the wait symbol is displayed.

Figure 2-88 Calibration wait pop-up

108
1 of 2
,
and the channel
The Zeroing pop- up is displayed.
Auto Zero A
and toggle the
Zero
softkey.
Zero
softkey - choosing
N1911A/1912A P-Series Power Meters User's Guide
to On.
Zero Both
Zero A Zero B
,
or