Power Meter Operation

Recorder Output

The rear panel Recorder Output connector produces a DC voltage that
corresponds to the power level in Watts of the channel, depending on the
measurement mode. This DC voltage ranges from 0 to +1 Vdc. The output
impedance is typically 1 kW . Channel and display offsets and duty cycle
have no effect on the Recorder Output.
For example, the Recorder Output can be used to;
• record swept measurements on an X- Y recorder
• level an output from a source using external leveling or
• monitor the output power on a strip chart recorder. A setup for
See
implemented in the recorder output.

Figure 2-20 Test Setup for Recording Swept Measurements

To access the "Recorder" menu, press
This menu allows you to switch the Recorder Output signal either on or
off. The
power level that you want to represent the 1 Vdc maximum and 0 Vdc
minimum output voltage of the Recorder Output.
86
recording swept measurements is shown in
Figure 2- 22
for details on which power meter functions are
Swept Source
SWEEP
RF OUT
OUT
X-AXIS (FREQUENCY)
and
Max Power
Device
Under Test
IN
OUT
X-Y Recorder
softkeys allow you to enter the input
Min Power
Figure 2-
20.
RECORDER
OUTPUT
Power Meter
CHANNEL A
INPUT
Y-AXIS (POWER)
,
,

Recorder Output

E4418B Power Meter User's Guide
.