Advertisement

7
Using 8480 Series Power Sensors
5 To change the frequency, press
6 Press
7 To confirm your choice, press the appropriate unit softkey.
8 Connect the power sensor to the signal to be measured.
9 The corrected measurement result is now displayed.
If the measurement frequency does not correspond directly to a frequency in the sensor
N O T E
calibration table, the power meter calculates the calibration factor using linear
interpolation.
If you enter a frequency outside the frequency range defined in the sensor calibration table,
the power meter uses the highest or lowest frequency point in the sensor calibration table
to set the calibration factor.
When
N O T E
sensor table identifier is displayed in the upper window. Also, pressing
displays the frequency you entered and calibration factor for each channel derived from the
selected sensor tables.

Figure 7-130Frequency/calibration table display

186
highlight the Frequency field.
to display the Frequency pop- up window. Use the numeric
keypad to enter the required value in the Frequency pop- up window.
Single Numeric
display mode is chosen, the frequency you entered and the
N1911A/1912A P-Series Power Meters User's Guide
and use the
and
keys to
,
Offset