Using 8480 Series Power Sensors

Sensor Calibration Tables

This section describes how to use sensor calibration tables. Sensor
calibration tables store the measurement calibration factors, for a power
sensor model or for a specific power sensor, in the power meter. They are
used to correct measurement results.
Use sensor calibration tables when you want to make power measurements
over a range of frequencies using one or more power sensors.
The P- Series power meters are capable of storing 20 sensor calibration
tables each containing up to 80 frequency points. The power meter is
supplied with a set of 9 predefined sensor calibration tables plus a "100%"
default table. The data in these tables is based on statistical averages for
a range of Agilent Technologies power sensors. Your own sensor will most
likely differ from the typical to some degree. If you require best accuracy,
create a custom table for each sensor you use as shown
in"Editing/Generating Sensor Calibration Tables"
To use calibration factor tables:
1 Select the sensor calibration table to be applied to a channel.
2 Zero and calibrate the power meter. The reference calibration factor
3 Specify the frequency of the signal you want to measure. The
4 Make the measurement.
184
used during the calibration is automatically set by the power meter
from the sensor calibration table.
calibration factor is automatically set by the power meter from the
sensor calibration table.
N1911A/1912A P-Series Power Meters User's Guide
on page 187.