Figure 4-108Trace Display With Trigger Delay Menu - Agilent Technologies N1911A User Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

Using E9320 E-Series Power Sensors
4
Select
Select Mrks Tgdel
Pressing
displays the gate markers or trigger
markers.
Markers
Mrks
When
is selected, Markers
and
indicate the start and
1
2
Marker
end points of the selected measurement gate. Pressing
toggles
between the two markers, the highlighted marker is the currently active
marker. Use the
and
keys to move the markers left or right across
the display.
The tables on the lower left of the screen show the time (
) and the
Time:
instantaneous power level (
) of the markers at their configured points.
Pow:
The table highlighted with the green border represents the active marker,
also highlighted in green. A negative time value indicates a measurement
before the trigger point. (See
Figure 4-
107).
Gate timing parameters are all related to your chosen trigger point. This may be different
N O T E
from the timing of the triggering event if you have configured a trigger delay.
Trigger Delay Value

Figure 4-108Trace display with trigger delay menu

TgDel
TgDel
When
is selected you can adjust the trigger delay. The
display changes to remove the gate markers and displays the trigger
marker(s). The
indicates when the trigger event occurs, whilst
shows
the delayed trigger point. When the two points coincide, only the
trigger is shown.
N1911A/1912A P-Series Power Meters User's Guide
143

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

1912aN1912a

Table of Contents