You can reduce the steps required to carry out the zero and calibration procedure as
N O T E
follows:
1 Connect the power sensor to the POWER REF output.
2 Press
Now set the sensor calibration factor for the frequency of the signal you
want to measure.
9 Check the current reference calibration factor setting by pressing
Does this setting match the value for the sensor? (The power sensor
reference calibration factor can normally be found above the calibration
factors table on the power sensor body.)
10 To change the settings, use the

Figure 8-139Calibration factor pop-up window

11 Press
12 Connect the power sensor to the signal to be measured.
13 The corrected measurement result is displayed.
N1911A/1912A P-Series Power Meters User's Guide
Zero + Cal
and
Zero + Cal
Zero + Cal A
,
Offset
. The value is displayed on the Cal Fac field.
,
value field and press
numeric keypad to enter the required value in the Cal Factor pop- up
window.
%
to complete the entry.
Using N8480 Series Power Sensors
. (For dual channel meters, press
Zero + Cal B
or
as required).
and
to display the Cal Factor pop- up. Use the
keys to highlight the Cal Fac
8